Skip to main content Skip to search Skip to main navigation

Temperature/pressure data logger EBI 12-TP231-EX

Product information "Temperature/pressure data logger EBI 12-TP231-EX"

Temperature/pressure data logger for process validation in the washer-disinfector, endoscope washer, and steam steriliser.

  • Data logger can be sterilised

  • 1 axially arranged external, rigid PT1000 temperature sensor

  • 1 internal piezoresistive pressure sensor with Luer-Lock connection

  • Real-time data transmission

  • Fast response time

  • Storage modes: Endless measurement, measurement upon start time, measure immediately until end of memory, start/stop measurement

  • Memory for up to 33440 measured values per measurement channel

  • Stainless steel housing

  • Factory calibration certificate: 0 °C, 60 °C, and 134 °C; 100 mbar and 3100 mbar at 25 °C, 3100 mbar at 134 °C

The logger is suitable for use in explosive gas atmospheres in zones 0/1/2 in accordance with Directive 2014/24 EU (ATEX Directive). ATEX-Type: II1G Ex ia IIC T4 Ga. Permissible ambient conditions in hazardous area: 0 to 85 °C, 100 Pa to 400 kPa.

colorpicker
Air pressure: 1 ... 4,000 mbar
Battery/rechargeable battery: 3.6 V, Lithium
Dimensions (ØxH): 48 x 32 mm
Electrochemical system: Lithium
Height: 32 mm
IP code: IP 68
Length sensor: 40 mm
Manufacturer Article Number: 1340-6655-EX
Material housing: Stainless steel
Max. air pressure: 4000 mbar
Max. measuring cycle: 24 h
Max. temperature: 150 °C
Max. voltage battery/rechargeable battery: 3.6 V
Measuring cycle: 250 ms ... 24 h
Min. air pressure: 1 mbar
Min. measuring cycle: 250 ms
Min. temperature: 0 °C
Resolution: 0.01 °C
Resolution air pressure: 1 mbar
Sensor (ØxL): 1.95 x 40 mm
Sensor Ø: 1.95 mm
Temperature: 0 ... 150 °C
Type: EBI 12-TP231-EX
Type of auxiliary energy: Battery
UNSPSC: 41112405
Weight: 120 g
Ø: 48 mm
Skip product gallery

Accessories

Interface Set SI 1100
Interface Set SI 1100

With the interface and the software, two data loggers can be programmed and the recorded data can be read out.Scope of supply: Interface IF 100, software Winlog.pro

Interface Set SI 2110

With the interface and the software, four data loggers can be programmed and the recorded data can be read out.Scope of supply: Interface IF 200, software Winlog.med, antenna

Interface Set SI 3200

With the interface and the software, four data loggers can be programmed and the recorded data can be read out.Scope of supply: Interface IF 200, software Winlog.validation, antenna

Interface Set SI 2100
Interface Set SI 2100

With the interface and the software, two data loggers can be programmed and the recorded data can be read out.Scope of supply: Interface IF 100, software Winlog.med

Interface Set SI 1200

With the interface and the software, four data loggers can be programmed and the recorded data can be read out.Scope of supply: Interface IF 200, software Winlog.pro, antenna