Skip to main content Skip to search Skip to main navigation

Temperature data logger EBI 12-T220-EX

Product information "Temperature data logger EBI 12-T220-EX"

Data logger for temperature measurement in steam sterilisers, washer-disinfectors, and endoscope washers during process validation.

  • 1 measuring channel

  • Wireless temperature measurement

  • Radial design with external, bendable PT1000 temperature sensor

  • Real-time data transmission

  • Fast response time

  • Storage modes: Endless measurement, measurement upon start time, measure immediately until end of memory, start/stop measurement

  • Memory for up to 100320 measured values

  • Stainless steel housing

  • Factory calibration certificate: -20 °C, 0 °C, 60 °C, and 134 °C

The logger is suitable for use in explosive gas atmospheres in zones 0/1/2 in accordance with Directive 2014/24 EU (ATEX Directive). ATEX-Type: II1G Ex ia IIC T4 Ga. Permissible ambient conditions in hazardous area: -40 to 85 °C, 100 Pa to 400 kPa.

colorpicker
Battery/rechargeable battery: 3.6 V, Lithium
Dimensions (ØxH): 48 x 24 mm
Disposable: No
Electrochemical system: Lithium
Height: 24 mm
IP code: IP 68
Length sensor: 250 mm
Manufacturer Article Number: 1340-6604-EX
Material housing: Stainless steel
Max. measuring cycle: 24 h
Max. temperature: 200 °C
Max. voltage battery/rechargeable battery: 3.6 V
Measuring cycle: 250 ms ... 24 h
Min. measuring cycle: 250 ms
Min. temperature: -200 °C
Resolution: 0.01 °C
Sensor (ØxL): 1.5 x 250 mm
Sensor Ø: 1.5 mm
Temperature: -200 ... 200 °C
Type: EBI 12-T220-EX
Type of auxiliary energy: Battery
UNSPSC: 41111979
Weight: 110 g
Ø: 48 mm
Skip product gallery

Accessories

New
Interface Set SI 2110
Interface Set SI 2110

With the interface and the software, four data loggers can be programmed and the recorded data can be read out.Scope of supply: Interface IF 200, software Winlog.med, antenna

Interface Set SI 2100
Interface Set SI 2100

With the interface and the software, two data loggers can be programmed and the recorded data can be read out.Scope of supply: Interface IF 100, software Winlog.med

Interface Set SI 1100
Interface Set SI 1100

With the interface and the software, two data loggers can be programmed and the recorded data can be read out.Scope of supply: Interface IF 100, software Winlog.pro

New
Interface Set SI 3200
Interface Set SI 3200

With the interface and the software, four data loggers can be programmed and the recorded data can be read out.Scope of supply: Interface IF 200, software Winlog.validation, antenna

New
Interface Set SI 1200
Interface Set SI 1200

With the interface and the software, four data loggers can be programmed and the recorded data can be read out.Scope of supply: Interface IF 200, software Winlog.pro, antenna