Microplate spectrophotometer with monochromator-based optical system. Wavelengths can be freely selected without changing filters. Suitable for spectral scanning, endpoint measurements, kinetic measurements, and scanning of well areas.

  • Measurements in 6 to 384 well microplates
  • Various quantitative and qualitative assay tests available
  • 7 different regressions to analyse data
  • Up to 6 wavelengths can be set in the endpoint/kinetic settings
  • Clear user interface

  • Monochromator-based UV/Vis absorbance


Microvolume measurements for quantification of nucleic acids possible with nano volume plate. Please order separately.


Detection system:
Photodiode
Wavelength range:
200 ... 999 nm
Dynamic range:
0 ... 4.0 OD
Dynamic accuracy:
0 ... 2 OD ±1 %
Dynamic repeatability:
0 ... 2 OD ±1 %
Dynamic linearity:
0 ... 2 OD ±1 %
Light source:
Xenon lamp
Depth: 303 mm
Dimensions (WxDxH): 333 x 303 x 245 mm
Height: 245 mm
Manufacturer Article Number: L-IN-A1
Power supply: 100 ... 240 V, 50/60 Hz
Type: INNO
Weight: 7.5 kg
Width: 333 mm
Skip product gallery

Other customers also bought

Microplate spectrophotometer INNO-M
Microplate spectrophotometer with monochromator-based optical system. Wavelengths can be freely selected without changing filters. Suitable for spectral scanning, endpoint measurements, kinetic measurements, and scanning of well areas.Measurements in 6 to 384 well microplatesVarious quantitative and qualitative assay tests available7 different regressions to analyse dataUp to 6 wavelengths can be set in the endpoint/kinetic settingsClear user interfaceMonochromator-based UV/Vis absorbanceWith luminescence detectionMicrovolume measurements for quantification of nucleic acids possible with nano volume plate. Please order separately.Detection system:Photodiode (absorbance), photomultiplier (luminescence)Wavelength range:200 ... 999 nm (absorbance), 300 ... 700 nm (luminescence)Dynamic range:0 ... 4.0 ODDynamic accuracy:0 ... 2 OD ±1 %Dynamic repeatability:0 ... 2 OD ±1 %Dynamic linearity:0 ... 2 OD ±1 %Light source:Xenon lamp

€0.00*
Skip product gallery

Accessories

Nano volume plate Nano-VC for microplate spectrophotometer INNO
Plate for holding a 24-well microtitre plate with 2 µl sample volume. Suitable for nucleic acid quantification up to 2 ng/µl dsDNA. With slot for a 2.5 ml cuvette with 0.5 mm path length.

€0.00*
Absorbance test plate for microplate spectrophotometer INNO
The absorbance test plate is specifically designed to check the performance of the device by controlling the linearity, accuracy, and alignment of the device.

€0.00*